FDA-Device2020-06-17Class II

therascreen EGFR RGQ PCR Kit (24), Reference Number REF 870121

Qiagen Sciences LLC
Hazard

There is a risk for a false mutation positive result in rare cases resulting from a fluorescence artefact. In these rare cases, the controls pass correctly, while a fluorescence artefact in the mutation assay causes an incorrect valid mutation positive result.

Sold states
US Nationwide.
Affected count
530 kits within expiry
Manufactured in
19300 Germantown Rd, N/A, Germantown, MD, United States
Products
therascreen EGFR RGQ PCR Kit (24), Reference Number REF 870121