FDA-Device2013-10-09Class II
Lactate Membrane units for E7077 Electrode Radiometer Medical ApS. The intended use is in vitro testing of samples of whole blood.
Radiometer America Inc
Hazard
RADIOMETER became aware that some membranes may have enzyme residue on the outer membrane. The enzyme residue may cause an initial negative bias on the reported Lactate result upon replacement of the Lactate membrane.
Sold states
Nationwide Distribution, including the states of: AK, AL, AR, AZ, BM, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI and WV.
Affected count
13,632 boxes/4 membranes
Manufactured in
810 Sharon Dr, Westlake, OH, United States
Products
Lactate Membrane units for E7077 Electrode Radiometer Medical ApS. The intended use is in vitro testing of samples of whole blood.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2303-2013Don't want to check this manually?