FDA-Device2016-08-10Class II
MultiDiagnost Eleva; x-ray system. Product Codes: 708032, 708036
Philips Electronics North America Corporation
Hazard
The two bolts that attach the clamping plate to the lateral shaft may break, which could cause the c-arm to move laterally forward, slide out of its bearings, and sag
Sold states
Worldwide distribution. US nationwide, Canada, Algeria, ARGENTINA, Australia, Austria, Belgium, China, Colombia, Czech Republic, Denmark, Egypt, Estonia, Faroe Islands, Finland, France, French Guiana, Germany, Greece, Hong Kong, Indonesia, Ireland, Israel, Italy, Jersey, Jordan, Republic of Korea, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Netherlands, New Zealand, Norway, Oman, Philippines, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, TURKEY, Ukraine, United Arab Emirates, and United Kingdom.
Affected count
993 total
Manufactured in
3000 Minuteman Rd, Andover, MA, United States
Products
MultiDiagnost Eleva; x-ray system. Product Codes: 708032, 708036
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2307-2016Don't want to check this manually?