FDA-Device2012-09-12Class II
Toshiba Ultra-Short Magnetic Resonance Imaging System, EXCELART Vantage. MRT-1503/P5-Vantage MRI System. MR imaging.
Toshiba American Medical Systems Inc
Hazard
The firm recalled due to a potentail problem with the position of the hold in the lug teminals on the power cables connected to the terminal blocks on the gradient coil.
Sold states
US (nationwide) distribution including the states of Ohio and Texas.
Affected count
3 units
Manufactured in
2441 Michelle Dr, P.O. Box 2068, Tustin, CA, United States
Products
Toshiba Ultra-Short Magnetic Resonance Imaging System, EXCELART Vantage. MRT-1503/P5-Vantage MRI System. MR imaging.
Official notice