FDA-Device2017-06-14Class II
Vitality Spinal Fixation System, Vitality T27 Final Drivers and Vitality Torque Limiting Handle 90 in-lbs Model Number(s): 07.02066.001, 07.02053.001, 730M0017
Zimmer Biomet Spine Inc.
Hazard
Zimmer Biomet is conducting a medical device recall for Vitality T27 Final Drivers and Vitality Torque Limiting Handle 90 in-lbs because when applying torque to tighten the closure top the driver has the potential to create a torsional spring which can cause damage and the potential for the values of torque to go out of specification.
Sold states
Nationwide, military/VA/govt and foreign consignee (No Canadian consignees)
Affected count
497 units
Manufactured in
10225 Westmoor Dr, N/A, Westminster, CO, United States
Products
Vitality Spinal Fixation System, Vitality T27 Final Drivers and Vitality Torque Limiting Handle 90 in-lbs Model Number(s): 07.02066.001, 07.02053.001, 730M0017
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2318-2017Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief