FDA-Device2015-08-05Class II

ACUSON S2000 and ACUSON S3000 ultrasound systems with the Virtual Touch IQ option; Model numbers: 10041461, 10440017 -S2000 system 10441730 -S3000 system VTIQ Option- 10439521 and 10439522 Radiology: The ultrasound imaging systems are intended for the following applications: Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Transcranial, OB/GYN, Cardiac, Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.

Siemens Medical Solutions USA, Inc.
Hazard

Potential measurement error on ACUSON S Family ultrasound system. When repositioning the Virtual Touch IQ region of interest from the original (default) location, the lateral position of the measurement tool may not align with the lateral position of the shear velocity data.

Sold states
Worldwide Distribution - US Nationwide and the countries of Algeria, Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Croatia, Czech Republic, Denmark, Dominican Rep., France, Germany, Greece, Hong Kong, India, Indonesia, Israel, Italy, Japan, Kenya, Malaysia, Mexico, Nepal, Netherlands, New Zealand, Norway, P.R. China, Philippines, Poland, Portugal, Republic Korea, Romania, Russian Fed., Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, U.A.E., Ukraine, United Kingdom.
Affected count
629 units
Manufactured in
685 E Middlefield Rd, N/A, Mountain View, CA, United States
Products
ACUSON S2000 and ACUSON S3000 ultrasound systems with the Virtual Touch IQ option; Model numbers: 10041461, 10440017 -S2000 system 10441730 -S3000 system VTIQ Option- 10439521 and 10439522 Radiology: The ultrasound imaging systems are intended for the following applications: Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Transcranial, OB/GYN, Cardiac, Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2321-2015

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief