FDA-Device2018-07-04Class II
Promote, Sterile EO, Model #/ Part #: 3207-30/60010740-207, 60010740-407; 3207-36/60010743-207, 60010743-407; 3213-36/50022126-001, 60013879-307, 60013879-407, 60013879-707; CD3211-36/100006984,100012548, 100012550, 100016680, 100031621, 100035618, 100035619, 100042487, 100047194; CD3211-36Q/100004091, 100004093,100019784, 100035620, 100042486, 100047054; CD3215-36/100006995, 100010246, 100010247, 100010249, 100010250, 100012569, 100012570, 100034981, 100046577, 100046802, 60019065;
St Jude Medical Inc.
Hazard
The firm is releasing new firmware for high voltage implantable cardiac devices (ICD) and cardiac resynchronization therapy defibrillators (CRT-D) intended to strengthen security by providing an additional layer of protection against unauthorized device access. For older generation devices not capable of accepting the firmware, the firm is advising customers of the option to permanently disable the RF communication capability in these devices.
Sold states
U.S. Nationwide distribution
Affected count
37485
Manufactured in
15900 Valley View Ct, Sylmar, CA, United States
Products
Promote, Sterile EO, Model #/ Part #: 3207-30/60010740-207, 60010740-407; 3207-36/60010743-207, 60010743-407; 3213-36/50022126-001, 60013879-307, 60013879-407, 60013879-707; CD3211-36/100006984,100012548, 100012550, 100016680, 100031621, 100035618, 100035619, 100042487, 100047194; CD3211-36Q/100004091, 100004093,100019784, 100035620, 100042486, 100047054; CD3215-36/100006995, 100010246, 100010247, 100010249, 100010250, 100012569, 100012570, 100034981, 100046577, 100046802, 60019065; CD3215-36Q/100004059, 100004060, 100004061, 100023304, 100030590, 100033678, 100042492, 100046803, 100046906
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2322-2018Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief