FDA-Device2025-08-27Class I

ENDOTAK RELIANCE G/SG, transvenous defibrillation lead, implantable cardioverter defibrillator (non-CRT); Model Nos. 0160, 0161, 0162, 0163, 0164, 0165, 0166, 0167, 0170, 0171, o182, 0173, 0174, 0175, 0176, 0177, 0180, 0182, 0183, 0184, 0185, 0186, 0187

Boston Scientific Corporation
Hazard

Affected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or need for cardiac resuscitation.

Sold states
Worldwide Distribution. US nationwide. International distribution worldwide.
Affected count
247,702 US; 344,670 OUS
Manufactured in
4100 Hamline Ave N, N/A, Saint Paul, MN, United States
Products
ENDOTAK RELIANCE G/SG, transvenous defibrillation lead, implantable cardioverter defibrillator (non-CRT); Model Nos. 0160, 0161, 0162, 0163, 0164, 0165, 0166, 0167, 0170, 0171, o182, 0173, 0174, 0175, 0176, 0177, 0180, 0182, 0183, 0184, 0185, 0186, 0187

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2325-2025