FDA-Device2012-09-19Class II

Siemens Uroskop Access. Fluoroscopic, image intensified x-ray system

Siemens Medical Solutions USA, Inc
Hazard

Siemens has discovered a possible hazard to patients or hospital personnel when tilting the Uroskop Access system starting with serial number 6076. When tilting the table on the Uroskop Access table, the motor may become disconnected and the entire table assembly may fall on the floor.

Sold states
Nationwide distribution: USA including states of: IN, MI, MS, TN and TX.
Affected count
5
Manufactured in
51 Valley Stream Pkwy, N/A, Malvern, PA, United States
Products
Siemens Uroskop Access. Fluoroscopic, image intensified x-ray system

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2341-2012