FDA-Device2012-09-19Class II
Siemens Uroskop Access. Fluoroscopic, image intensified x-ray system
Siemens Medical Solutions USA, Inc
Hazard
Siemens has discovered a possible hazard to patients or hospital personnel when tilting the Uroskop Access system starting with serial number 6076. When tilting the table on the Uroskop Access table, the motor may become disconnected and the entire table assembly may fall on the floor.
Sold states
Nationwide distribution: USA including states of: IN, MI, MS, TN and TX.
Affected count
5
Manufactured in
51 Valley Stream Pkwy, N/A, Malvern, PA, United States
Products
Siemens Uroskop Access. Fluoroscopic, image intensified x-ray system
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2341-2012Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief