FDA-Device2025-09-03Class I
Battery Operated infusion pumps. t:slim X2 G5, Replacement, Refurbished, REF: 1000911; t:slim X2 GS Classic, Replacement, Refurbished, REF: 1003808; t: slim X2, Basal-IQ, mg/dl, REF: 1004219; Pump, t:slim X2, Clinical Use Only, REF: 1004484; Pump, t:slim X2, Control-IQ, REF: 1005611; Pump, t:slim X2, Control-IQ, mmoUL, Refurbished Replacement, REF: 1005613; Pump, t:slim X2, Control-IQ, mg/dl, REF: 1005615; Pump, t:slim X2, Control-IQ, mg/dl, Refurbished Replacement, REF: 1005617; Pump,
Tandem Diabetes Care, Inc.
Hazard
Insulin pumps manufactured with speakers from Revision A and B may malfunction ("Malfunction 16") which prevents audible alerts and causes the pumps to become inoperable and stop infusing insulin which could result in a hyperglycemic event, which may lead to injury and/or hospitalization.
Sold states
Worldwide distribution. US nationwide including Puerto Rico, Canada, Czech Republic, Slovakia, Spain, United Kingdom, Belgium, Luxembourg, Australia, France, Italy, New Zealand, Finland, Sweden, Denmark, Norway, Saudi Arabia, Germany, Netherlands, Ireland, Portugal, South Africa, Switzerland
Affected count
171,750
Manufactured in
12400 High Bluff Dr, San Diego, CA, United States
Products
Battery Operated infusion pumps. t:slim X2 G5, Replacement, Refurbished, REF: 1000911; t:slim X2 GS Classic, Replacement, Refurbished, REF: 1003808; t: slim X2, Basal-IQ, mg/dl, REF: 1004219; Pump, t:slim X2, Clinical Use Only, REF: 1004484; Pump, t:slim X2, Control-IQ, REF: 1005611; Pump, t:slim X2, Control-IQ, mmoUL, Refurbished Replacement, REF: 1005613; Pump, t:slim X2, Control-IQ, mg/dl, REF: 1005615; Pump, t:slim X2, Control-IQ, mg/dl, Refurbished Replacement, REF: 1005617; Pump, t:slim X2, Basal-IQ 6.4, Refurbished Replacement, REF: 1006379; Pump, t:slim X2, Control-IQ 7.4, REF: 1006402; Pump, t:slim X2, Control-IQ 7.4, Replacement, REF: 1006406; Pump, t:slim X2, Control-IQ 7.4, REF: 1006408; Pump, t:slim X2, Basal-IQ, , REF: 1006419; Pump, t:slim X2, Basal-IQ, mg/dL, REF: 1006420; Pump, t:slim X2, Basal-IQ, mmoUL, Refurbished Replacement, REF: 1006535; Pump, t:slim X2, Basal-IQ, mg/dL, Refurbished Replacement, REF: 1006537; Pump, t:slim X2, Control-IQ, mg/dl , Replacement, REF: 1008255; Pump, t:slim X2, Control-IQ, mmoUL, Replacement, REF: 1008256; Pump, t:slim X2, Basal-IQ, mmoUL, Replacement, REF: 1008257; Pump, t:slim X2, Basal-IQ, mg/dl ,Replacement, REF: 1008258; Pump, t:slim X2, Basal-IQ, REF: 1010004; Pump, t:slim X2, Basal-IQ, Replacement, REF: 1010005; Pump, t:slim X2, Basal-IQ, Medicare, REF: 1010006; Pump, t:slim X2, Basal-IQ, Medicare, REF: 1010007; Pump, t:slim X2, Control-IQ, REF: 1010009; Pump, t:slim X2, Control-IQ Replacement, REF: 1010010; Pump, t:slim X2, Control-IQ, Medicare, REF: 1010011; Pump, t:slim X2, Control-IQ, Refurbished Replacement, REF: 1010012; Pump, t:slim X2, Control-IQ, REF: 1012867; Pump, t:slim X2, Control-IQ, Medicare, REF: 1012868; Pump, t:slim X2, Control-IQ, Replacement, REF: 1012869; Pump, t:slim X2, Control-IQ 7.8, REF: 1013195; Pump, t:slim X2, Control-IQ 7.8, Medicare, REF: 1013196; Pump, t :slim X2, Control-IQ 7.8, Replacement, REF: 1013197; Pump, t:slim X2, Control-IQ 7.8, Refurbished Replacement, REF: 1013198
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2343-2025Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief