FDA-Device2021-09-08Class II
Baxter PrismaFlex, REF numbers: a) 106696 (M 60 SET) b) 106697 (M 100 SET) Expansion: c) 107144 (TPE2000 SET) d) 115307 (M 150 SET) e) 115309 (ST 100 SET) f) 955467 (M 150 SET) g) 955468 (ST 100 SET) h) 955595 (M 100 SET) i) 955596 (ST 150 SET) Used with Prismaflex or PrisMax control units for continuous fluid management and renal replacement therapies.
Hazard
Products labeled as sterile were distributed, but may not have been sterilized. " Baxter confirmed Steril Milano observed nonconformances that were not reported to Baxter. Data was manipulated to appear to be conforming
Sold states
US Nationwide distribution.
Affected count
64275 units
Manufactured in
1 Baxter Pkwy, N/A, Deerfield, IL, United States
Products
Baxter PrismaFlex, REF numbers: a) 106696 (M 60 SET) b) 106697 (M 100 SET) Expansion: c) 107144 (TPE2000 SET) d) 115307 (M 150 SET) e) 115309 (ST 100 SET) f) 955467 (M 150 SET) g) 955468 (ST 100 SET) h) 955595 (M 100 SET) i) 955596 (ST 150 SET) Used with Prismaflex or PrisMax control units for continuous fluid management and renal replacement therapies.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2378-2021More Baxter Healthcare Corporation recalls
- FDA-Device2026-07-08Hillrom VOLARA System P.CIRCUIT KIT AC recalled for nebulizer cup leakage risk
- FDA-Device2026-07-08Baxter operating table head positioning adapter recalled for loosening risk
- FDA-Device2026-07-08Hillrom VOLARA System P.CIRCUIT 5KIT AC recalled for nebulizer cup leakage risk
- FDA-Device2026-07-08Hillrom VOLARA System P.CIRCUIT KIT recalled for nebulizer cup leakage risk
- FDA-Device2026-07-08Hillrom Volara P.Circuit 5Kit breathing device recalled for nebulizer cup leakage risk
Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief