FDA-Device2021-09-08Class II
Baxter prismaflex, HF 1000 PAES membrane, REF 107140 Used with Prismaflex or PrisMax control units for continuous fluid management and renal replacement therapies.
Baxter Healthcare Corporation
Hazard
Products labeled as sterile were distributed, but may not have been sterilized. Baxter confirmed Steril Milano observed nonconformances that were not reported to Baxter Data was manipulated to appear to be conforming
Sold states
US Nationwide distribution.
Affected count
1976 units
Manufactured in
1 Baxter Pkwy, N/A, Deerfield, IL, United States
Products
Baxter prismaflex, HF 1000 PAES membrane, REF 107140 Used with Prismaflex or PrisMax control units for continuous fluid management and renal replacement therapies.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2379-2021Don't want to check this manually?