FDA-Device2021-09-08Class II

Baxter prismaflex, HF 1000 PAES membrane, REF 107140 Used with Prismaflex or PrisMax control units for continuous fluid management and renal replacement therapies.

Baxter Healthcare Corporation
Hazard

Products labeled as sterile were distributed, but may not have been sterilized. Baxter confirmed Steril Milano observed nonconformances that were not reported to Baxter Data was manipulated to appear to be conforming

Sold states
US Nationwide distribution.
Affected count
1976 units
Manufactured in
1 Baxter Pkwy, N/A, Deerfield, IL, United States
Products
Baxter prismaflex, HF 1000 PAES membrane, REF 107140 Used with Prismaflex or PrisMax control units for continuous fluid management and renal replacement therapies.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2379-2021

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief