FDA-Device2018-07-18Class II

da Vinci Xi EndoWrist Stapler 45 Reload Blue, Model 48645B-03, 12 reloads/box.

Intuitive Surgical, Inc.
Hazard

Intuitive Surgical, Inc. (ISI) has recently become aware of an issue with the da Vinci Xi Stapler 45 Blue and Green Reloads that may be missing a component of the knife mechanism. Without this component, the reload knife will not move and cut tissue when the stapler is fired.

Sold states
U.S., foreign and government accounts. Foreign: Australia, Austria, Belgium, Brazil, Canada, China (Hong Kong), Czech Republic, Denmark, Finland, France, Germany, Ireland, Italy, Japan, Luxembourg, Netherlands, Norway, Qatar, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom.
Affected count
7337
Manufactured in
1266 Kifer Rd Bldg 101, Sunnyvale, CA, United States
Products
da Vinci Xi EndoWrist Stapler 45 Reload Blue, Model 48645B-03, 12 reloads/box.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2380-2018