FDA-Device2023-08-16Class II
MR 7700 Magnetic Resonance (MR) system, Product Numbers 782120 and 782130 (Upgrade to MR 7700)
Philips North America
Hazard
The system operator is able to bypass the SmokeDetector Interlock system locking mechanism after a smoke detection by power cycling the system.
Sold states
US Nationwide. Global Distribution.
Affected count
640 total
Manufactured in
222 Jacobs St, N/A, Cambridge, MA, United States
Products
MR 7700 Magnetic Resonance (MR) system, Product Numbers 782120 and 782130 (Upgrade to MR 7700)
Official notice