FDA-Device2020-07-01Class II
MEVION S250i, MEVION S250; Proton Radiation Treatment System
Mevion Medical Systems, Inc.
Hazard
Couch Correction moves sent after using a 3D CT scan are partially lost if the previous correction sent or recorded had some of the Couch Correction checkboxes turned off
Sold states
Distribution to US in MO, NJ, OH, OK, FL, and DC, and the Netherlands
Affected count
8 units
Manufactured in
300 Foster St, N/A, Littleton, MA, United States
Products
MEVION S250i, MEVION S250; Proton Radiation Treatment System
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2406-2020Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief