FDA-Device2024-07-31Class II
Dxl 9000 Access Immunoassay Analyzer, Catalog #C11137.
Hazard
The analyzers were built with nonconforming USB cables made with brass contacts instead of specified gold contacts. Brass contacts may corrode over time, leading to loss of internal instrument communication and delayed patient results.
Sold states
Worldwide - US Nationwide distribution in the states of FL, IN, KS, NH, PA, and TX. The countries of Australia, Austria, Belgium, Brazil, Croatia, Czechia, Egypt, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, Spain, Switzerland, Taiwan, Province of China, and United Kingdom of Great Britain and Northern Ireland.
Affected count
143 devices
Manufactured in
1000 Lake Hazeltine Dr, Chaska, MN, United States
Products
Dxl 9000 Access Immunoassay Analyzer, Catalog #C11137.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2414-2024More Beckman Coulter, Inc. recalls
- FDA-Device2026-04-15Beckman Coulter MicroScan Neg MIC 3J antibiotic susceptibility panels recalled for manufacturing defects
- FDA-Device2026-04-15MicroScan Neg Multidrug Resistant MIC 1 REF C32368 UDI-DI code: 15099590720568 Intended Use- For use with MicroScan Dried Gram Negative MIC/Combo Panels and Dried Gram Negative Breakpoint Combo Panels. MicroScan panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of aerobic and facultatively anaerobic gram-negative bacilli.
- FDA-Device2026-04-01Beckman Coulter MicroScan Neg MIC 56 panels recalled for potential well contamination
Don't want to check this manually?