FDA-Device2024-08-07Class II
SlideView VS200 slide viewer,
EVIDENT SCIENTIFIC INC
Hazard
A broken fuse inside the laser that was caused by a short circuit at PIN 4 of the DOOR CTRL. This short circuit was caused by a badly positioned cable that made contact with the housing/frame and caused a short circuit and destroyed the fuse in the laser. This caused the shutter to remain in its most recent state; either in an open or closed position. Once the fuse was destroyed the shutter could not be operated.
Sold states
US
Affected count
N/A
Manufactured in
48 Woerd Ave, N/A, Waltham, MA, United States
Products
SlideView VS200 slide viewer,
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2416-2024Don't want to check this manually?