FDA-Device2015-08-26Class II

Siemens Symbia S Series system, a diagnostic analog or digital detector based planar (2- dimensional) nuclear medicine (NM) imaging system installed and used in only one location within an imaging facility or in a mobile/ transportable imaging environment.

Siemens Medical Solutions USA, Inc.
Hazard

Fuses with the improper amperage rating were installed in the Symbia S, Symbia T Series and Symbia lntevo systems. These fuses, in conjunction with a specific failure mode of the Power Controller assembly, could lead to the overheating of the assembly.

Sold states
Worldwide Distribution - USA including AL, AK, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY, District of Columbia, Puerto Rico and Internationally to Algeria, Argentina , Australia , Austria , Bangladesh , Belgium , Brazil , Bulgaria , Canada , Chile , China, Colombia , Costa Rica , Croatia , Czech Republic , Denmark , Ecuador , Egypt , Estonia, Finland , France , Germany , Hong Kong , Hungary , Iceland , India , Indonesia , Iran , Ireland , Israel , Italy , Japan , Jordan , Kazakhstan , Korea, Kuwait , Latvia , Libya , Lithuania , Luxembourg , Malaysia , Mexico , Morocco , Myanmar , Netherlands , New Zealand , Norway , Oman , Pakistan , Peru , Philippines , Poland , Portugal , Qatar , Romania , Russia, Saudi Arabia , Singapore , Slovakia , Slovenia , South Africa , Spain , Sweden , Switzerland , Taiwan , Tanzania, Thailand , Tunisia , Turkey , United Arab Emirates , Ukraine , United Kingdom , Venezuela , Vietnam
Affected count
1183 Systems
Manufactured in
2501 Barrington Rd, Hoffman Estates, IL, United States
Products
Siemens Symbia S Series system, a diagnostic analog or digital detector based planar (2- dimensional) nuclear medicine (NM) imaging system installed and used in only one location within an imaging facility or in a mobile/ transportable imaging environment.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2435-2015