Nihon Kohden NKV-550 Ventilator, REF: NKV-550-U, CE 0344, UDI: (01) 00843685100012 The Nihon Kohden NKV-550 Series Ventilator System is intended to provide continuous ventilation for adult, pediatric and neonatal patients who require invasive or noninvasive respiratory support.
NIHON KOHDEN ORANGEMED, INC
Hazard
Potential defective gas inlet pressure regulator within ventilator which could result in gas releasing noise from the bottom of the ventilator, a High Air/O2 Inlet Pressure alarm, a High/Low O2 alarm, or could not pass device Check prior to use.
Sold states
Worldwide Distribution: US (nationwide) to states of: AZ, CA, NY, MA, VT, LA, DC, PA, FL, UT and OUS (countries) of: Brazil, Japan, Ireland, Poland, Scotland, Germany, Kazakhstan, Spain, Slovenia, Germany/Bulgaria, Russia, India, Palestine, Libya, Jordan Kuwait, Lebanon, Syria, Turkey, South Africa, United Arab Emirates, Mexico, Colombia, Peru, Costa Rica, Barbados, Ecuador, Bahamas, Bolivia, St Kitts & Nevis, St. Martin, Chile, Thailand, Laos, Philippines, Malaysia, Nepal and United Kingdom.
Affected count
1,999 units
Manufactured in
1800 E Wilshire Ave, N/A, Santa Ana, CA, United States
Products
Nihon Kohden NKV-550 Ventilator, REF: NKV-550-U, CE 0344, UDI: (01) 00843685100012 The Nihon Kohden NKV-550 Series Ventilator System is intended to provide continuous ventilation for adult, pediatric and neonatal patients who require invasive or noninvasive respiratory support.