FDA-Device2020-07-08Class II

Control Panel REF - 60-02-15: LivaNova SCP Pump Control Panel. REF 60-02-15, IPX2, RxOnly, UDI: 04033817900986

LivaNova Deutschland GmbH
Hazard

The SCP control Panel rotary knob used to control rotary speed of the pump in the SCP system used as a centrifugal pump during cardiopulmonary bypass may be difficult or impossible to rotate as a result of premature wear of the angle encoder. The loss of function of the angle encoder during bypass may lead to ischemia, stroke, organ damage, arterial tissue injury or reperfusion syndrome.

Sold states
US: FL, GA, IN, KY, MI, MO, NY, OH, PA, SC, TN, TX, VA OUS: Canada - Uzbekistan - Turkey - Tunisia - Taiwan - South Africa - Serbia - Saudi Arabia - Russia - Poland - Mexico - Korea(Rep. of) - India - Hong Kong - Colombia - China - Brazil - Bahrain - Argentina - Algeria - Italy - Germany
Affected count
24
Manufactured in
Lindberghstr. 25, N/A, Munich, N/A, Germany
Products
Control Panel REF - 60-02-15: LivaNova SCP Pump Control Panel. REF 60-02-15, IPX2, RxOnly, UDI: 04033817900986

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2440-2020