FDA-Device2019-09-11Class II

AngioDynamics VenaCure EVLT NeverTouch Procedure Kit, 45CM NEVERTOUCH FRS .018 PROCEDURE KIT PG, Catalog Number 11403012

Angiodynamics, Inc.
Hazard

The affected product is properly labelled with an expiration date of March 31, 2022; but the RFID programmed expiration date is April-2014 which results in an error message when connected to the laser source and prevents the product from being used.

Sold states
The products were distributed to the following US states: DC, IL, IN, MI, NJ, TN, TX, and WI.
Affected count
109
Manufactured in
603 Queensbury Ave, N/A, Queensbury, NY, United States
Products
AngioDynamics VenaCure EVLT NeverTouch Procedure Kit, 45CM NEVERTOUCH FRS .018 PROCEDURE KIT PG, Catalog Number 11403012

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2444-2019

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief