FDA-Device2021-09-22Class II
syngo.via RT Image Suite with syngo.via VB30, VB40, VB50 and VB60- Intended for Radiation Therapy Treatment Material Number: 10496180
Siemens Medical Solutions USA, Inc
Hazard
Potential for data loss when using the contouring or patient marking workflow when editing a structure set after saving and reopening a study, resulting in an incomplete structure set sent to the treatment planning system (TPS).
Sold states
US Nationwide distribution.
Affected count
126 units
Manufactured in
40 Liberty Blvd, N/A, Malvern, PA, United States
Products
syngo.via RT Image Suite with syngo.via VB30, VB40, VB50 and VB60- Intended for Radiation Therapy Treatment Material Number: 10496180
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2445-2021Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief