FDA-Device2021-09-22Class II

syngo.via RT Image Suite with syngo.via VB30, VB40, VB50 and VB60- Intended for Radiation Therapy Treatment Material Number: 10496180

Siemens Medical Solutions USA, Inc
Hazard

Potential for data loss when using the contouring or patient marking workflow when editing a structure set after saving and reopening a study, resulting in an incomplete structure set sent to the treatment planning system (TPS).

Sold states
US Nationwide distribution.
Affected count
126 units
Manufactured in
40 Liberty Blvd, N/A, Malvern, PA, United States
Products
syngo.via RT Image Suite with syngo.via VB30, VB40, VB50 and VB60- Intended for Radiation Therapy Treatment Material Number: 10496180