FDA-Device2024-08-21Class I

Replacement Battery List Number SUB0000594 and SUB0000864 found in Plum A+ & Plum A+3 Infusion Systems, Pump List Numbers: 11005, 11971, 12391, 12618, 20678, 20679, 20792, 60529, 12348, 11973. OUS units marketed with the following descriptions: Plum A+ Mednet TM Firmware New 13.6 Span, Plum A+3, PLUM A+3 INFUSION PUMPS and Plum A+3 Wireless Refurb, PLUM A+, PLUM A+ Infusion System, PlumA+3IV Pump Multi 1' Set

ICU Medical Inc
Hazard

Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the battery lifecycle than expected.

Sold states
Worldwide Distribution: US (nationwide) and OUS (foreign) countries of: Australia, Canada, France, Guadeloupe, Guatemala, Ireland, Italy, Oman, Paraguay, Qatar, Spain, Saudi Arabia, South Africa, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Uruguay.
Affected count
14,744 units
Manufactured in
600 N Field Dr, Lake Forest, IL, United States
Products
Replacement Battery List Number SUB0000594 and SUB0000864 found in Plum A+ & Plum A+3 Infusion Systems, Pump List Numbers: 11005, 11971, 12391, 12618, 20678, 20679, 20792, 60529, 12348, 11973. OUS units marketed with the following descriptions: Plum A+ Mednet TM Firmware New 13.6 Span, Plum A+3, PLUM A+3 INFUSION PUMPS and Plum A+3 Wireless Refurb, PLUM A+, PLUM A+ Infusion System, PlumA+3IV Pump Multi 1' Set

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2448-2024

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief