FDA-Device2015-09-02Class II
Dimension Vista B¿2-Microglobulin Flex Reagent Cartridge (B2MIC); Product Usage: The B2MIC method is an in vitro diagnostic test for the quantitative measurement of B¿2-Microglobulin in human serum, heparinized plasma, EDTA plasma and urine on the Dimension Vista System. Measurements of B¿2-Microglobulin aid in the diagnosis of active rheumatoid arthritis and kidney disease.
Siemens Healthcare Diagnostics, Inc.
Hazard
customers are receiving a high number of Abnormal Assay Errors or are not able to calibrate B2MIC when using Flex reagent cartridge B2MIC various lots. In failure mode, there is the potential for falsely depressed beta2-Microglobulin results on the Dimension Vista system due to assay reagent contamination.
Sold states
US Nationwide Distribution in the states of TN, AZ, CT, Fl, CA, NY and WI
Affected count
Total 1655 units
Manufactured in
500 Gbc Dr Ms 514, PO BOX 6101, Newark, DE, United States
Products
Dimension Vista B¿2-Microglobulin Flex Reagent Cartridge (B2MIC); Product Usage: The B2MIC method is an in vitro diagnostic test for the quantitative measurement of B¿2-Microglobulin in human serum, heparinized plasma, EDTA plasma and urine on the Dimension Vista System. Measurements of B¿2-Microglobulin aid in the diagnosis of active rheumatoid arthritis and kidney disease.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2450-2015Don't want to check this manually?