FDA-Device2019-09-18Class I
Natrelle 133 Tissue Expanders (also named style 133 BIOSPAN Tissue Expander and Mcghan Magna-Site Tissue Expander) BioCell Full Height Variable Projection with Magna-Fider Xact & 21G Needle Infusion Set, Sterile, for the following style numbers: STYLE 133FV: 133FV-11, 133FV-12, 133FV-13, 133FV-14, 133FV-15, 133FV-16; STYLE 133MV: 133MV-11, 133MV-12, 133MV-13, 133MV-14, 133MV-15, 133MV-16; STYLE 133LV: 133LV-11, 133LV-12, 133LV-13, 133LV-14, 133LV-15, 133LV-16; S
Allergan PLC
Hazard
The action was initiated following notification by the U.S. FDA of their recently updated, global safety information concerning the higher incidence of anaplastic large cell lymphoma (BIA-ALCL) in patients who have textured breast implants.
Sold states
Worldwide Distribution and US Nationwide All fifty (50) US States, the US Virgin Islands and Puerto Rico. (AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY). OUS: Bermuda, Canada, Chile, Japan, Korea, Thailand, Taiwan and Vietnam.
Affected count
4,026,287 Breast Implants and Tissue Expanders Combined in total
Manufactured in
2525 Dupont Dr, Irvine, CA, United States
Products
Natrelle 133 Tissue Expanders (also named style 133 BIOSPAN Tissue Expander and Mcghan Magna-Site Tissue Expander) BioCell Full Height Variable Projection with Magna-Fider Xact & 21G Needle Infusion Set, Sterile, for the following style numbers: STYLE 133FV: 133FV-11, 133FV-12, 133FV-13, 133FV-14, 133FV-15, 133FV-16; STYLE 133MV: 133MV-11, 133MV-12, 133MV-13, 133MV-14, 133MV-15, 133MV-16; STYLE 133LV: 133LV-11, 133LV-12, 133LV-13, 133LV-14, 133LV-15, 133LV-16; STYLE 133FX: 133FX-11 133FX-12 133FX-13 133FX-14 133FX-15 133FX-16 STYLE 133MX: 133MX-11, 133MX-12, 133MX-13, 133MX-14, 133MX-15, 133MX-16; STYLE 133SX: 133SX-11, 133SX-12, 133SX-13, 133SX-14, 133SX-15, 133SX-16; STYLE 133SV: 133SV-11, 133SV-12, 133SV-13, 133SV-14, 133SV-15, 133SV-16; These products may have been sold under the McGhan, Inamed or NATRELLE brands. Product Usage: intended for temporary subcutaneous implantation and require periodic, incremental inflation with sterile saline for injection until the desired amount of tissue is developed.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2455-2019Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief