FDA-Device2025-09-10Class I

IDM-ARLITE-R. Electrical wheelchair component.

mo-Vis BVBA
Hazard

Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.

Sold states
US Nationwide distribution in the state of TX.
Affected count
25 units
Manufactured in
Biebuyckstraat 15D, N/A, Nevele, N/A, Belgium
Products
IDM-ARLITE-R. Electrical wheelchair component.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2459-2025