FDA-Device2025-09-03Class II
AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) CYSTO, Model/Item Number ASCY24A; 2) CYSTO PACK, Model/Item Number BUCY78E; 3) CYSTO PACK, Model/Item Number LMCP22T; 4) TUR BASIN PACK, Model/Item Number LMTU31T; 5) CYSTO, Model/Item Number SACY80S; 6) UROLOGY PACK, Model/Item Number SPUR64F.
American Contract Systems Inc
Hazard
During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
Sold states
US distribution to states of: IL and LA
Affected count
1127 units
Manufactured in
85 Shaffer Park Dr, Tiffin, OH, United States
Products
AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) CYSTO, Model/Item Number ASCY24A; 2) CYSTO PACK, Model/Item Number BUCY78E; 3) CYSTO PACK, Model/Item Number LMCP22T; 4) TUR BASIN PACK, Model/Item Number LMTU31T; 5) CYSTO, Model/Item Number SACY80S; 6) UROLOGY PACK, Model/Item Number SPUR64F.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2466-2025Don't want to check this manually?