FDA-Device2025-09-03Class II

AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) LAPAROSCOPY PACK, MODEL/ITEM NUMBER EVLP61I; 2) PELVISCOPY, MODEL/ITEM NUMBER EVPV29G.

American Contract Systems Inc
Hazard

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Sold states
US distribution to states of: IL and LA
Affected count
1094 units
Manufactured in
85 Shaffer Park Dr, Tiffin, OH, United States
Products
AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) LAPAROSCOPY PACK, MODEL/ITEM NUMBER EVLP61I; 2) PELVISCOPY, MODEL/ITEM NUMBER EVPV29G.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2470-2025