FDA-Device2023-09-06Class II

Atellica CH Total Bilirubin_2 -In vitro diagnostic use in the quantitative determination of total bilirubin in human serum and plasma (lithium heparin) of adults and neonates Test Code: TBil_2 Siemens Material Number (SMN): 11097531

Siemens Healthcare Diagnostics, Inc.
Hazard

Potential for reagent carryover on the Atellica CH 930 resulting in a positive bias that could impact quality control (QC), patient samples, and calibrator results with from Atellica¿ CH Total Bilirubin_2 (TBil_2), Atellica¿CH LDL Cholesterol (LDLC), Atellica¿ CH Gamma-Glutamyl Transferase (GGT) and Atellica¿CH HDL Cholesterol (HDLC)

Sold states
US Nationwide - Worldwide Distribution Foreign: Taiwan Uruguay Vietnam
Affected count
50148 units
Manufactured in
511 Benedict Ave, N/A, Tarrytown, NY, United States
Products
Atellica CH Total Bilirubin_2 -In vitro diagnostic use in the quantitative determination of total bilirubin in human serum and plasma (lithium heparin) of adults and neonates Test Code: TBil_2 Siemens Material Number (SMN): 11097531

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2471-2023