FDA-Device2012-10-10Class II
12/14 Neck Taper 00789503201 VerSys Femoral Head Provisional 32mm -3.5 00789503202 VerSys Femoral Head Provisional 32mm +0 00789503203 VerSys Femoral Head Provisional 32mm +3.5 00789503204 VerSys Femoral Head Provisional 32mm +7 00789503205 VerSys Femoral Head Provisional 32mm +10.5 The provisional heads replicate final femoral implant geometry and are used to trial range of motion, joint stability, and determine appropriate femoral head size and offset during total hip arthroplasty.
Zimmer, Inc.
Hazard
Complaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in some cases alleging the fit was too loose and in other cases, alleging the fit was too tight. Intraoperative dislodging of the provisional head and the provisional head being left in-vivo after separating from mating comp
Sold states
Worldwide Distribution - USA (nationwide) and internationally to Canada, Mexico, South America, Europe, Puerto Rico, China, and Korea.
Affected count
231,641 all lots, all sizes
Manufactured in
345 E Main St, N/A, Warsaw, IN, United States
Products
12/14 Neck Taper 00789503201 VerSys Femoral Head Provisional 32mm -3.5 00789503202 VerSys Femoral Head Provisional 32mm +0 00789503203 VerSys Femoral Head Provisional 32mm +3.5 00789503204 VerSys Femoral Head Provisional 32mm +7 00789503205 VerSys Femoral Head Provisional 32mm +10.5 The provisional heads replicate final femoral implant geometry and are used to trial range of motion, joint stability, and determine appropriate femoral head size and offset during total hip arthroplasty.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2476-2012Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief