FDA-Device2015-09-02Class II

OPD-Scan III Refractive Power/Corneal Analyzer Opthalmic Software versions 1.00.08 1.05.07 1.10.01, 1.01.02 1.06.02 1.11.02, 1.02.01 1.07.01 1.12.03, 1.03.02 1.08.01 1.13.01, 1.04.03 1.09.01. Opthalmic: The OPD-Scan III is a diagnostic instrument that is indicated for use for Mapping of refractive error distribution of the eye by measurement and analysis of spherical power, cylindrical power, and cylinder axis.

Nidek Inc
Hazard

Software bug was found where there was no difference in Total and Corneal high-order aberrations, but differences were found in Internal high-order aberrations.

Sold states
Nationwide Distribution.
Affected count
3836
Manufactured in
47651 Westinghouse Dr, N/A, Fremont, CA, United States
Products
OPD-Scan III Refractive Power/Corneal Analyzer Opthalmic Software versions 1.00.08 1.05.07 1.10.01, 1.01.02 1.06.02 1.11.02, 1.02.01 1.07.01 1.12.03, 1.03.02 1.08.01 1.13.01, 1.04.03 1.09.01. Opthalmic: The OPD-Scan III is a diagnostic instrument that is indicated for use for Mapping of refractive error distribution of the eye by measurement and analysis of spherical power, cylindrical power, and cylinder axis.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2477-2015