FDA-Device2015-09-16Class I

OmniPod¿ Insulin Management System. Only the OmniPod¿ device is affected. Catalog Number(s): POD-ZXP420 (US). Intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin.

Insulet Corporation
Hazard

OmniPods¿ (Pods) have a higher rate of failure causing: Cannula fails to deploy/retract causing insulin to be pumped or the audible alarm is displayed on the PDM and pod will not deliver insulin

Sold states
Worldwide Distribution: US (Nationwide) and country of: Switzerland.
Affected count
24829 boxes Expanded: 26,230.9 boxes
Manufactured in
600 Technology Park Dr Ste 200, N/A, Billerica, MA, United States
Products
OmniPod¿ Insulin Management System. Only the OmniPod¿ device is affected. Catalog Number(s): POD-ZXP420 (US). Intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2484-2015