FDA-Device2025-09-10Class II
Superion IDS Kit Superion Indirect Decompression System 8MM, VF Implant 101-9808 00884662000529 Superion Indirect Decompression System 10MM, VF Implant 101-9810 00884662000536 Superion Indirect Decompression System 12MM, VF Implant 101-9812 00884662000543 Superion Indirect Decompression System 14MM, VF Implant 101-9814 00884662000550 Superion Indirect Decompression System 16MM, VF Implant 101-9816 00884662000567 Superion Indirect Decompression System Kit, VF Instrument
Boston Scientific Neuromodulation Corporation
Hazard
As a result of a comprehensive product performance review it was determine that the device and instruments are not consistently meeting expected performance levels. Issues include breakage of the instruments and implant/device.
Sold states
U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY.
Affected count
70,516
Manufactured in
25155 Rye Canyon Loop, N/A, Valencia, CA, United States
Products
Superion IDS Kit Superion Indirect Decompression System 8MM, VF Implant 101-9808 00884662000529 Superion Indirect Decompression System 10MM, VF Implant 101-9810 00884662000536 Superion Indirect Decompression System 12MM, VF Implant 101-9812 00884662000543 Superion Indirect Decompression System 14MM, VF Implant 101-9814 00884662000550 Superion Indirect Decompression System 16MM, VF Implant 101-9816 00884662000567 Superion Indirect Decompression System Kit, VF Instrument 102-9800 00884662000574 The Vertiflex Procedure Instrument Platform, VF Instrument 140-9800 00884662000611 The Superion IDS Kit includes a set of proprietary instruments to deliver the Superion Implant in a minimally invasive manner. Instruments specifically designed for implanting the Superion Implant are sterile, single-use disposable instruments, consisting of a Dilator Assembly, Cannula Assembly, Interspinous Gauge, Inserter, Reamer, and Driver.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2485-2025Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.