FDA-Device2012-10-10Class II
Cemented Tibial Drill Product Usage: The cemented tibial drill is used to prepare the tibial bone prior to the impaction of the applicable stemmed tibial broach. The cemented tibial drill has engraved lines that are size specific to guide the user as to the depth to insert the drill into the tibial bone.
Zimmer, Inc.
Hazard
Zimmer is initiating a recall of the Cemented Tibial Drill for the Zimmer Persona Personalized Knee System due to three reports being received as of August 29. 2012 for tibial cortex perforation during preparation of the tibial bone. The investigation into the reports that have been received found that the perforations are occurring while using the Cemented Tibial Drill. The perforations have o
Sold states
Worldwide Distribution - US Nationwide including the states of: AZ, CA, CO,FL, MN, IL, NJ, NY, OH, OR and the countries of Austrailia, Utd. Arab Emir., France, Germany, India, and Italy.
Affected count
60
Manufactured in
345 E Main St, N/A, Warsaw, IN, United States
Products
Cemented Tibial Drill Product Usage: The cemented tibial drill is used to prepare the tibial bone prior to the impaction of the applicable stemmed tibial broach. The cemented tibial drill has engraved lines that are size specific to guide the user as to the depth to insert the drill into the tibial bone.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2493-2012Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.