FDA-Device2012-10-10Class II

Philips Digital Diagnost with patient carriage Model: 712050 with serial number range from 00 000 00 to 09 000 543. Product Usage: Stationary radiographic System

Philips Healthcare Inc.
Hazard

Hook does not securely hold the footplate in vertical position, causing the footplate to disengage and may fall on patient or operator

Sold states
Worldwide Distribution - US Nationwide and the countries of :Australia, Austria, Bahrain, Belgium, Canada, Chile, China, Czech Republic, Denmark, Finland, , France, Germany, Ireland, Italy, Japan, Netherlands, New Zealand, Norway, Poland, Qatar, Romania, Russia, Saudi Arabia, Sweden, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan. Thailand, Turkey,United Arab Emirates and United Kingdom.
Affected count
536 units
Manufactured in
3000 Minuteman Road, N/A, Andover, MA, United States
Products
Philips Digital Diagnost with patient carriage Model: 712050 with serial number range from 00 000 00 to 09 000 543. Product Usage: Stationary radiographic System

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2494-2012