FDA-Device2024-08-14Class II
TruDi NAV Suction, 0 Degrees; Model/Catalog Number: TDNS000Z. TruDi" NAV Suction Instruments are intended for use with the TruDi" Navigation System during surgical procedures in ENT medicine and skull base surgery
Integra LifeSciences Corp.
Hazard
Specific lots of TruDi NAV Suction Instruments 0 Degrees were incorrectly calibrated which when used with the TruDi Navigation System, may result in a discrepancy between the actual tip location and the visualized location displayed on the system.
Sold states
The devices are only sold in the US. The following states have received the product: AL, AZ CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, PA, SC, SD, TN, TX, VA, VT, WI.
Affected count
265 units
Manufactured in
1100 Campus Rd, Princeton, NJ, United States
Products
TruDi NAV Suction, 0 Degrees; Model/Catalog Number: TDNS000Z. TruDi" NAV Suction Instruments are intended for use with the TruDi" Navigation System during surgical procedures in ENT medicine and skull base surgery