FDA-Device2024-08-14Class II
Baxter PST 500 U, Precision Surgical Table, Part number (product code) 4080300
Baxter Healthcare Corporation
Hazard
Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction for the PST 500 U Precision Surgical Table due to a potential issue with loose spring pins in the spindle drives. This can cause the tabletop to unexpectedly tilt or move at any time, even without active use of the surgical table. This could potentially result in unintentional patient movement during surgical procedures and/or preparation for surgical procedures, including transport.
Sold states
Worldwide distribution - US Nationwide and the country of Canada.
Affected count
400 units
Manufactured in
1 Baxter Pkwy, Deerfield, IL, United States
Products
Baxter PST 500 U, Precision Surgical Table, Part number (product code) 4080300
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2512-2024Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief