Baxter PST 500 U, Precision Surgical Table, Part number (product code) 4080300

Baxter Healthcare Corporation
Hazard

Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction for the PST 500 U Precision Surgical Table due to a potential issue with loose spring pins in the spindle drives. This can cause the tabletop to unexpectedly tilt or move at any time, even without active use of the surgical table. This could potentially result in unintentional patient movement during surgical procedures and/or preparation for surgical procedures, including transport.

Sold states
Worldwide distribution - US Nationwide and the country of Canada.
Affected count
400 units
Manufactured in
1 Baxter Pkwy, Deerfield, IL, United States
Products
Baxter PST 500 U, Precision Surgical Table, Part number (product code) 4080300

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2512-2024