FDA-Device2020-07-15Class II

uCT 760 System, Computed Tomography X-ray System

Shanghai United Imaging Healthcare Co., Ltd.
Hazard

Two issues were identified with the computed tomography x-ray system including an intermittent issue which may affect image data leading to artifacts; and intermittent scout scanning interruption due to occasional angle signal drift. This may result in the need for rescanning of the patient with the patient receiving an additional dose of radiation and possible use of additional contrast media.

Sold states
Distributed to two customers. Both customers located in Houston, Texas.
Affected count
5 units
Manufactured in
No. 2258 Chengbei Rd, Jiading Ind., Shanghai, China
Products
uCT 760 System, Computed Tomography X-ray System

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2522-2020