FDA-Device2023-09-27Class I
Hamilton-T1, REF: 161006, Intensive Care and Transport Ventilator
Hamilton Medical, Inc.
Hazard
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
Sold states
Distribution US nationwide, including Puerto Rico. OUS: Not provided
Affected count
76 units
Manufactured in
4655 Aircenter Cir, Reno, NV, United States
Products
Hamilton-T1, REF: 161006, Intensive Care and Transport Ventilator
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2526-2023Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief