FDA-Device2020-07-15Class II

Curaplex Ambu King LTS-D Supraglottic Airway Kit with ET Tube Part Number: KLTSD404K Curaplex Ambu¿ King LTS-D" Supraglottic Airway Kit with ET Tube (hereafter the Kit which contains all of the components necessary to successfully obtain an open patient airway with an Ambu¿ King LTS-D" Supraglottic Airway.

Bound Tree Medical
Hazard

Kit contained an incorrect catheter-tip syringe not compatible with the luer lock connector of the included airway. As a result, may be unable to inflate the balloons of the supraglottic airway to create a secure seal in the patient s airway as intended by the kit

Sold states
Nationwide Distribution
Affected count
167 kits
Manufactured in
5000 Tuttle Crossing Blvd, Dublin, OH, United States
Products
Curaplex Ambu King LTS-D Supraglottic Airway Kit with ET Tube Part Number: KLTSD404K Curaplex Ambu¿ King LTS-D" Supraglottic Airway Kit with ET Tube (hereafter the Kit which contains all of the components necessary to successfully obtain an open patient airway with an Ambu¿ King LTS-D" Supraglottic Airway.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2532-2020

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief