FDA-Device2020-07-15Class II
Curaplex Ambu King LTS-D Supraglottic Airway Kit with ET Tube Part Number: KLTSD404K Curaplex Ambu¿ King LTS-D" Supraglottic Airway Kit with ET Tube (hereafter the Kit which contains all of the components necessary to successfully obtain an open patient airway with an Ambu¿ King LTS-D" Supraglottic Airway.
Bound Tree Medical
Hazard
Kit contained an incorrect catheter-tip syringe not compatible with the luer lock connector of the included airway. As a result, may be unable to inflate the balloons of the supraglottic airway to create a secure seal in the patient s airway as intended by the kit
Sold states
Nationwide Distribution
Affected count
167 kits
Manufactured in
5000 Tuttle Crossing Blvd, Dublin, OH, United States
Products
Curaplex Ambu King LTS-D Supraglottic Airway Kit with ET Tube Part Number: KLTSD404K Curaplex Ambu¿ King LTS-D" Supraglottic Airway Kit with ET Tube (hereafter the Kit which contains all of the components necessary to successfully obtain an open patient airway with an Ambu¿ King LTS-D" Supraglottic Airway.