FDA-Device2014-09-03Class II
Medtronic MiniMed Paradigm Insulin Infusion Pumps, Paradigm models: MMT-511, MMT-512, MMT-712, MMT-712E, MMT-515, MMT-715, MMT-522, MMT-522K, MMT-722, MMT-722K, MMT-523, MMT-523K, MMT-723, MMT-723K, MMT-554, and MMT-754. MiniMed 530G: MMT-551, and MMT-751 Paradigm insulin infusion pumps are intended for use to deliver basal and bolus insulin to persons with diabetes.
Medtronic MiniMed Inc.
Hazard
Medtronic is recalling the Medtronic MiniMed Paradigm Insulin Infusion Pumps because they have received reports that users have accidentally programmed the pump to deliver the maximum bolus amount.
Sold states
Worldwide Distribution-Nationwide in US
Affected count
559,374 units total (444,374 units in US)
Manufactured in
18000 Devonshire St, N/A, Northridge, CA, United States
Products
Medtronic MiniMed Paradigm Insulin Infusion Pumps, Paradigm models: MMT-511, MMT-512, MMT-712, MMT-712E, MMT-515, MMT-715, MMT-522, MMT-522K, MMT-722, MMT-722K, MMT-523, MMT-523K, MMT-723, MMT-723K, MMT-554, and MMT-754. MiniMed 530G: MMT-551, and MMT-751 Paradigm insulin infusion pumps are intended for use to deliver basal and bolus insulin to persons with diabetes.
Official notice