FDA-Device2014-09-03Class II

Trauma Tubes and Plates: TUBE & SCP PLT 90DX150LG TUBE & SCP PLT 90DX205LG TUBE & SCP PLT 90DX255LG TUBE & SCP PLT 95DX150LG TUBE & SCP PLT 95DX205LG TUBE & SCP PLT 95DX255LG TUBE&SCP PLT 130DX8H TUBE&SCP PLT 130DX10H TUBE&SCP PLT 130DX12H TUBE&SCP PLT 130DX14H TUBE&SCP PLT 135DX8H TUBE&SCP PLT 135DX10H TUBE&SCP PLT 135DX12H TUBE&SCP PLT 135DX14H TUBE&SCP PLT 140DX8H TUBE&SCP PLT 140DX10H TUBE&SCP PLT 140DX12H TUBE&SCP PLT 140DX14H TUBE&SCP PLT 145DX8H TUBE&SCP PLT 145DX10H

Zimmer, Inc.
Hazard

Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.

Sold states
Worldwide Distribution.
Affected count
972,125 units
Manufactured in
1800 W Center St, N/A, Warsaw, IN, United States
Products
Trauma Tubes and Plates: TUBE & SCP PLT 90DX150LG TUBE & SCP PLT 90DX205LG TUBE & SCP PLT 90DX255LG TUBE & SCP PLT 95DX150LG TUBE & SCP PLT 95DX205LG TUBE & SCP PLT 95DX255LG TUBE&SCP PLT 130DX8H TUBE&SCP PLT 130DX10H TUBE&SCP PLT 130DX12H TUBE&SCP PLT 130DX14H TUBE&SCP PLT 135DX8H TUBE&SCP PLT 135DX10H TUBE&SCP PLT 135DX12H TUBE&SCP PLT 135DX14H TUBE&SCP PLT 140DX8H TUBE&SCP PLT 140DX10H TUBE&SCP PLT 140DX12H TUBE&SCP PLT 140DX14H TUBE&SCP PLT 145DX8H TUBE&SCP PLT 145DX10H TUBE&SCP PLT 145DX12H TUBE&SCP PLT 145DX14H TUBE&SCP PLT 150DX6H TUBE&SCP PLT 150DX8H TUBE&SCP PLT 150DX10H TUBE&SCP PLT 150DX12H TUBE&SCP PLT 150DX14H VERSA-FX II SC TUBE PLT 90DX6H VERSA-FX II SC TUBE PLT 90DX8H VERSA-FXII SC TUBE PLT 90DX10H VERSA-FXII SC TUBE PLT 90DX12H VERSA-FXII SC TUBE PLT 90DX14H VERSA-FXII SC TUBE PLT 95DX6H VERSA-FXII SC TUBE PLT 95DX8H VERSA-FXII SC TUBE PLT 95DX10H VERSA-FXII SC TUBE PLT 95DX12H VERSA-FXII SC TUBE PLT 95DX14H VERSA-FX KEYLESS 90D X 8HOLE VERSA-FX KEYLESS 90D X 11HOLE VERSA-FX KEYLESS 95D X 8HOLE VERSA-FX KEYLESS 95D X 11HOLE VERSA-FX KEYLESS 95D X 14HOLE VERSA-FX II STD TUBE 130DX8H VERSA-FX II STD TUBE 130DX10H VERSA-FX II STD TUBE 130DX12H VERSA-FX II STD TUBE 130DX14H VERSA-FX II STD TUBE 135DX8H VERSA-FX II STD TUBE 135DX10H VERSA-FX II STD TUBE 135DX12H VERSA-FX II STD TUBE 135DX14H VERSA-FX II STD TUBE 140DX8H VERSA-FX II STD TUBE 140DX10H VERSA-FX II STD TUBE 140DX12H VERSA-FX II STD TUBE 140DX14H VERSA-FX II STD TUBE 145DX8H VERSA-FX II STD TUBE 145DX10H VERSA-FX II STD TUBE 145DX12H VERSA-FX II STD TUBE 145DX14H VERSA-FX II STD TUBE 150DX8H VERSA-FX II STD TUBE 150DX10H VERSA-FX II STD TUBE 150DX12H VERSA-FX II STD TUBE 150DX14H VERSA-FX II KEYLESS TB 130DX8H VERSA-FX II KEYLESS TB 135DX8H VERSA-FX II KEYLESS TB 140DX8H VERSA-FX II KEYLESS TB 145DX8H VERSA-FX II KEYLESS TB 150DX8H VERSA-FX KEYLESS 90D X 8HOLE VERSA-FX KEYLESS 90D X 11HOLE VERSA-FX KEYLESS 95D X 8HOLE VERSA-FX KEYLESS 95D X 11HOLE VERSA-FX KEYLESS 95D X 14HOLE

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2550-2014

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief