FDA-Device2018-08-01Class II
cobas m 511 integrated hematology analyzer, Model Number 07261691190 Product Usage: The cobas m 511 integrated hematology analyzer is a quantitative, automated analyzer with cell locating capability. It is intended for in vitro diagnostic use by a skilled operator in the clinical laboratory. The system prepares a stained microscope slide from EDTA anticoagulated whole blood. It utilizes computer imaging to count the formed elements of blood and provide an image-based assessment of cell morph
Roche Diagnostics Hematology
Hazard
Discrepant results have been reported in patients with severe microcytic anemia (e.g.iron deficiency, thalassemia) and the following parameters are affected: RBC (red blood cell count), HGB (hemoglobin concentration), MCH (mean corpuscular hemoglobin), HCT (hematocrit), and MCV (mean corpuscular volume).
Sold states
No distribution in the United States. The devices were distributed to the following foreign countries: Austria, France, Germany, Hong Kong, Japan, Netherlands, Pakistan, Singapore, South Korea, Switzerland, Thailand, and Vietnam.
Affected count
17
Manufactured in
69 Milk St Ste 120, N/A, Westborough, MA, United States
Products
cobas m 511 integrated hematology analyzer, Model Number 07261691190 Product Usage: The cobas m 511 integrated hematology analyzer is a quantitative, automated analyzer with cell locating capability. It is intended for in vitro diagnostic use by a skilled operator in the clinical laboratory. The system prepares a stained microscope slide from EDTA anticoagulated whole blood. It utilizes computer imaging to count the formed elements of blood and provide an image-based assessment of cell morphology, which may be reviewed by the operator, and also allows for manual classification of unclassified cells. The instrument reports the following parameters: RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, %NRBC, #NRBC, WBC, %NEUT, #NEUT, %LYMPH, #LYMPH, %MONO, #MONO, %EO, #EO, %BASO, #BASO, PLT, MPV, %RET, #RET, HGB-RET.