FDA-Device2018-08-08Class II
Discovery IQ 2.0 20cm, 4 ring, Model Number 5432539-22
GE Healthcare, LLC
Hazard
Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).
Sold states
Nationwide distribution to AL, CA, CO, FL, IL, IN, MA, MD, MI, MS, NC, NJ, NY,OH, OK, SC, TN, TX, VA.
Affected count
42 devices total
Manufactured in
3000 N Grandview Blvd, N/A, Waukesha, WI, United States
Products
Discovery IQ 2.0 20cm, 4 ring, Model Number 5432539-22
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2578-2018Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.