FDA-Device2024-08-21Class II
HeartMate 3 System Controllers provided within the following HeartMate 3 Left Ventricular Assist System (LVAS) Implant Kits or distributed separately: HeartMate 3 LVAS Implant Kit, US, Model: 106524US; HeartMate 3 LVAS Implant Kit, OUS, Model: 106524INT; HeartMate 3 System Controller, US, Model: 106531US; HeartMate 3 System Controller, OUS, Model: 106531INT; HeartMate 3 System Controller Low Flow 2.0, Global, Model: 106531LF2
Thoratec LLC
Hazard
Left ventricular assist system controller UI membrane/screen may lift along the edge of the controller housing, and if controller is exposed to fluid, ingress may occur, which may damage printed circuit board assemblies, which could result in unexpected and false positive alarms, loss of visual alarms, loss of user interface(use of buttons, loss of visual alarms,etc.), loss of power, and pump stop
Sold states
US: MN, PA, FL, IL, NY, DE, AR, NJ, NC, GA, WI, AZ, TN, MO, TX, MA, OH, VA, NE, CA, OK, CO, DC, IN, MI, CT, UT, IA, MD, OR, NM, ME, SC, KY, LA, WA, NV, HI, AL, MS, KS, WV. OUS: Algeria, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia & Herzegovina, Brazil, Bulgaria, Canada, Cayman Islands, Chile, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Iran, Islamic Republic of, Iraq, Ireland, Israel, Italy, Korea, Kuwait, Lebanon, Lithuania, Macedonia, Morocco, Netherlands, New Zealand, Norway, Oman, Palestine, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uzbekistan
Affected count
694
Manufactured in
6035 Stoneridge Dr, Pleasanton, CA, United States
Products
HeartMate 3 System Controllers provided within the following HeartMate 3 Left Ventricular Assist System (LVAS) Implant Kits or distributed separately: HeartMate 3 LVAS Implant Kit, US, Model: 106524US; HeartMate 3 LVAS Implant Kit, OUS, Model: 106524INT; HeartMate 3 System Controller, US, Model: 106531US; HeartMate 3 System Controller, OUS, Model: 106531INT; HeartMate 3 System Controller Low Flow 2.0, Global, Model: 106531LF2
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2583-2024Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.