FDA-Device2017-06-28Class II

Proteus 235

Ion Beam Applications S.A.
Hazard

It has been observed during testing that there could be clinically non-negligible amount of dark current (a well-known physics effect observed in cyclotron) in Pencil Beam Scanning (PBS) treatment mode when system is not irradiating between layers.

Sold states
Worldwide Distribution - US Distribution to the states of : MA, FL, VA, PA, IL, NJ, WA, TN, LA and TX., and to the countries of: Germany, South Korea, France, Czech Republic, Italy, Poland and Sweden.
Affected count
18 systems ( 10 in the US)
Manufactured in
Chemin du Cyclotron, 3, Louvain La Neuve, Belgium
Products
Proteus 235

Official notice