FDA-Device2025-10-08Class I

Reprocessed ACUSON AcuNav Ultrasound Catheter for intracardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart. For imaging guidance only, not treatment delivery, during cardiac interventional percutaneous procedures. Medline Item Numbers 10135910RH 10135936RH

Medline Industries, LP
Hazard

Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.

Sold states
Domestic US distribution nationwide. International distribution pending. No international distribution.
Affected count
7913 eaches
Manufactured in
3 Lakes Dr, N/A, Northfield, IL, United States
Products
Reprocessed ACUSON AcuNav Ultrasound Catheter for intracardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart. For imaging guidance only, not treatment delivery, during cardiac interventional percutaneous procedures. Medline Item Numbers 10135910RH 10135936RH

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2614-2025

Don't want to check this manually?