FDA-Device2021-10-06Class II

Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model Nos. 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85

Datascope Corp.
Hazard

The Cardiosave IABP may unexpectedly shut down when the device is running on AC power, only one battery is installed in the IABP, and the battery is physically removed while the battery is being charged. This occurs during a very specific set of conditions.

Sold states
Nationwide domestic distribution. Foreign distribution worldwide.
Affected count
3969 total (US); 3550 total (OUS)
Manufactured in
1300 Macarthur Blvd, N/A, Mahwah, NJ, United States
Products
Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model Nos. 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2616-2021

Don't want to check this manually?