FDA-Device2014-09-17Class II

The BD MAX MRSA Assay, catalogue #442953. An automated qualitative in vitro diagnostic test for the direct detection of Methicillin-resistant Staphylococcus aureus (MRSA) DNA from nasal swabs in patie.

Becton Dickinson & Co.
Hazard

The performance of the assay no longer conforms to this statement in the Analytical Specificity section of the Package Insert: One hundred-eleven (111) out of 111 MSSA strains tested at extremely high concentrations (> 106 CFU/swab), produced negative results with the BD MAX MRSA Assay. One of the samples tested with each of the CAP surveys MRS5-A 2014 and MRS5-B 2014 yielded false positive results.

Sold states
US Distribution including the states of : CA, FL, HI, IL, IN, MD, MI, MN, MT, NC, ND, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA and WI.
Affected count
585,048 tests
Manufactured in
BD Diagnostic Systems, 7 Loveton Circle, Sparks, MD, United States
Products
The BD MAX MRSA Assay, catalogue #442953. An automated qualitative in vitro diagnostic test for the direct detection of Methicillin-resistant Staphylococcus aureus (MRSA) DNA from nasal swabs in patie.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2618-2014