FDA-Device2023-10-04Class II

exoplan version 3.1 Rijeka Software -A medical software, intended to support the pre-operative planning of dental implants using the visualization of the implant placement within images of the patents anatomy.

Exocad GmbH
Hazard

A software library filtering error has been discovered which occurs when users are using the Step-by-Step Full Drill Protocol export functionality of exoplan 3.1 Rijeka. When a guided surgery treatment approach is selected, along with a Step-by-Step Full Drill Protocol export implant library, exoplan software does not filter out unsupported sleeve height positions for a particular sub-full drill protocol and instead shows ALL possible sleeve height positions for all sub-full drill protocols. The use of an improper sleeve height position in a surgical guide could result in patient injury

Sold states
Worldwide distribution - US Nationwide and the country of Canada.
Affected count
96 licenses
Manufactured in
Julius-Reiber-Str. 37, N/A, Darmstadt, N/A, Germany
Products
exoplan version 3.1 Rijeka Software -A medical software, intended to support the pre-operative planning of dental implants using the visualization of the implant placement within images of the patents anatomy.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2633-2023